Organic sulphur compounds (thiosemicarbaxide and dithioxamide) are determined in the range 5.00-30.0 m ml-'. Electrolysis is done between two platinum electrodes in a closed vapour-generation system. Various parameters which affect the procedure are investigated in detail. Other sulphur-containing c
Continuous-flow molecular emission cavity analysis for organic sulphur compounds by alkaline hydrolysis
โ Scribed by N. Grekas; A.C. Calokerinos
- Book ID
- 104100626
- Publisher
- Elsevier Science
- Year
- 1988
- Tongue
- English
- Weight
- 460 KB
- Volume
- 204
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
โฆ Synopsis
Alkaline hydrolysis of various orgamc sulphur compounds (thioacetamide, thiosemicarbazide, thiodiacetic acid and dithiooxamide) in a continuous-flow analyzer produces sulphide which can be removed from the solution after acidification. Hydrogen sulphide evolved is measured by molecular emission cavity analysis. Samples can be analyzed at a rate of 30 h-' with a relative error of l-2%.