The potential of the ion trap mass spectrometer to characterize isomeric long-chain conjugated mono-or diethylenic epoxides was investigated using the multi-stage tandem mass spectrometric (MS n ) facilities of the instrument. In the proposed approach, mass-selected reagent ions obtained from common
Comparison of Tandem Mass Spectrometric Techniques Using Ion-trap and Hybrid Mass Spectrometers for Studies of the Degradation of a Herbicide by Biofilm
β Scribed by G. Kopf; J. V. Headley; K. M. Peru; R. C. Johnson; M. H. Soni; R. G. Cooks
- Publisher
- John Wiley and Sons
- Year
- 1997
- Tongue
- English
- Weight
- 156 KB
- Volume
- 11
- Category
- Article
- ISSN
- 0951-4198
No coin nor oath required. For personal study only.
β¦ Synopsis
The herbicide, methyl 2-4-(2,4-dichlorophenoxy)phenoxy proprionate (diclofop-methyl), and three of its degradation products are identified in biofilm using tandem mass spectrometry (MS/MS). Sub-samples of a biofilm cultured in a flow-through system, with the herbicide diclofop-methyl in the nutritive solution, were analyzed using two ion-trap mass spectrometers and the data were compared with those obtained using a hybrid instrument of EBEQ geometry. In order to achieve high throughput, the samples were introduced directly into the ion sources with no further sample preparation. Confirmations were based on data taken for authentic standards. The results indicate that data comparable to those obtained using the hybrid instrument could be achieved using ion-traps, despite the complex nature of the biofilm. This is true both of mass spectra and of product ion MS/MS spectra which were essential for identification of the individual compounds in these complex mixtures. For high-throughput analyses (20-50 runs daily), the quality of the MS/MS characterization was dependent on the background pressure and the source temperature.
π SIMILAR VOLUMES
Because of the lack of a UV chromophore and their much smaller abundances in comparison with the major component, the minor components in erythromycin estolate preparations are difficult to analyze by high performance liquid chromatography ultraviolet (HPLC-UV). Tentative assignment of the major and
A procedure is described for the determination of three characteristic b-agonists (clenbuterol, terbutalin and salbutamol) based on the formation of the corresponding protonated molecules and related collisional experiments. Quantification was carried out on selected collisional fragments and the re
The electrospray ionisation (ESI) of selected coumarin derivatives and their subsequent fragmentation using an ion trap mass spectrometer have been investigated. Sequential product ion fragmentation experiments (MS(n)) were performed in order to elucidate the degradation pathways for these compounds
A screening method was developed to specifically detect and identify isotopically labeled compounds automatically in real-time using a HPLC/ion trap mass spectrometer utilizing newly developed isotopic pattern recognition software. The ion trap mass spectrometer specifically detected isotopically la