## Purpose: To compare a free breathing navigator triggered single shot echoplanar imaging (ss epi) diffusion-weighted imaging (dwi) sequence with prospective acquisition correction (pace) with a breathhold (bh) dwi sequence for liver imaging. ## Materials and methods: Thirty-four patients were e
Comparison and reproducibility of ADC measurements in breathhold, respiratory triggered, and free-breathing diffusion-weighted MR imaging of the liver
✍ Scribed by Thomas C. Kwee; Taro Takahara; Dow-Mu Koh; Rutger A.J. Nievelstein; Peter R. Luijten
- Publisher
- John Wiley and Sons
- Year
- 2008
- Tongue
- English
- Weight
- 502 KB
- Volume
- 28
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose
To compare and determine the reproducibility of apparent diffusion coefficient (ADC) measurements of the normal liver parenchyma in breathhold, respiratory triggered, and free‐breathing diffusion‐weighted magnetic resonance imaging (DWI).
Materials and Methods
Eleven healthy volunteers underwent three series of DWI. Each DWI series consisted of one breathhold, one respiratory triggered, and two free‐breathing (thick and thin slice acquisition) scans of the liver, at b‐values of 0 and 500 s/mm^2^. ADCs of the liver parenchyma were compared by using nonparametric tests. Reproducibility was assessed by the Bland–Altman method.
Results
Mean ADCs (in 10^−3^ mm^2^/sec) in respiratory triggered DWI (2.07–2.27) were significantly higher than mean ADCs in breathhold DWI (1.57–1.62), thick slice free‐breathing DWI (1.62–1.65), and thin slice free‐breathing DWI (1.57–1.66) (P < 0.005). Ranges of mean difference in ADC measurement ± limits of agreement between two scans were −0.02–0.05 ± 0.16–0.24 in breathhold DWI, −0.14–0.20 ± 0.59–0.60 in respiratory triggered DWI, −0.03–0.03 ± 0.20–0.29 in thick slice free‐breathing DWI, and −0.01–0.09 ± 0.21–0.29 in thin slice free‐breathing DWI.
Conclusion
ADC measurements of the normal liver parenchyma in respiratory triggered DWI are significantly higher and less reproducible than in breathhold and free‐breathing DWI. J. Magn. Reson. Imaging 2008;28:1141–1148. © 2008 Wiley‐Liss, Inc.