Biosensors in flow-injection systems for biomedical analysis, process and environmental monitoring
β Scribed by Canh Tran-Minh
- Publisher
- John Wiley and Sons
- Year
- 1996
- Tongue
- English
- Weight
- 464 KB
- Volume
- 9
- Category
- Article
- ISSN
- 0952-3499
No coin nor oath required. For personal study only.
β¦ Synopsis
This paper presents the construction of various biosensors using thin-film layers incorporated in flow injection devices, providing automated systems for biomedical analysis, process and environmental monitoring. A urease sensor has been developed in conjunction with a flow injection system for the automatic determination of urea. Use of the spraying immobilization technique gives rise to a response time of a few seconds, which allows sample throughputs up to 200 h-'. With a penicillin biosensor adapted in an appropriate cell detection, on-line measurements of penicillin V in the fermentation broth are achieved during the whole fermentation process; the results are compared with the HPLC method. Linearity, sensitivity and reproducibility of the biosensor are studied with regards to sample dilution in a stirred flow detection cell to provide optimal operating conditions.
Measurements without any change in parameters are obtained during the whole fermentation process. Acetylcholinesterase sensors have been used in batch systems for the determination of pesticides, but they require large amounts of substrate. When those enzyme sensors are combined with flow injection systems, only small volumes (100 pl) of substrate are injected into the carrier stream and an automated system can be obtained for continuous control of water quality.
π SIMILAR VOLUMES
Screen printed ampemmetric glucose biosen~ors were constructed and optimised for use in a flow injection analysis (FIA) system. The sensors were fabricated using a catalytic metallised carbon based ink, allowing a working potential of + 350 mV (Ag/AgCI). Overall the sensors displayed a high degree o
An amperometric biosensor for determination of penicillin was developed. Penicillinase was immobilized by cross-linking with glutaraldehyde and bovine serum albumin, which was dropped on a platinum disc electrode. The principle of this electrode is based on the classical iodometric penicillin assay.
A flow-injection system with immobilized dehydrogenases, including sample pretreatments by means of a continuously working filtration unit for elimination of microorganisms and a dialyser, was applied to the parallel determination of glucose and ethanol during a baker's yeast fermentation. The syst
The preparation of cartridges with five different enzymes: glucose oxidase (GOD), pyranose oxidase (PyOD), alcohol oxidase ( AOD) , L-amino acid oxidase ( t-ASOD) and L-lactate-2-monooxigenase (LMO) immobilized on VA-Epoxy Biosynth carrier were optimized and standardized. The enzyme cartridges were