Gas chromatographic equipment and procedures are described for automated splitless injection of pseudo-static headspace samples collected externally onto a sorbent trap. The GC microprocessor controls, in sequence, carrier gas backflushing of the sorbent trap for water removal, splitless thermal des
Automated at-column injection into narrow bore capillary GC columns
✍ Scribed by Hagman, Gunnar ;Roeraade, Johan
- Publisher
- John Wiley and Sons
- Year
- 1991
- Tongue
- English
- Weight
- 283 KB
- Volume
- 14
- Category
- Article
- ISSN
- 0935-6304
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
An injector designed for automatic direct liquid injection into narrow bore capillary GC columns has been constructed and evaluated. The tip of the syringe needle is aligned with, and positioned close to, the column entrance in a small, pressurized cavity: when the sample is dispensed it is immediately forced into the column by the action of the surrounding carrier gas. A standard autosampler equipped with a standard stainless steel syringe needle was utilized for at‐column sample transfer into 100 μm i.d. columns.
RSD values for n‐alkanes were between 0.1 and 0.3% for relative area counts and approximately 1% for absolute area counts.