Asymmetric quadrature split birdcage coil for hyperpolarized 3He lung MRI at 1.5T

✍ Scribed by Nicola De Zanche; Navjeet Chhina; Kevin Teh; Chris Randell; Klaas P. Pruessmann; Jim M. Wild


Publisher
John Wiley and Sons
Year
2008
Tongue
English
Weight
777 KB
Volume
60
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

An asymmetric quadrature birdcage body coil for hyperpolarized (HP) ^3^He lung imaging at 1.5T is presented. The coil is designed to rest on top of the patient support and be used as a temporary insert in a clinical system. A two‐part construction facilitates patient access and the asymmetric design makes maximal use of available bore space to ensure comfort. Highly homogeneous, circularly polarized RF magnetic fields are produced at 48.5 MHz using a conformal mapping method for the geometrical design, combined with an algebraic method to calculate the individual capacitance values on the birdcage coil's ladder network. Efficiency and isolation from the system's proton body coil are ensured by an integrated RF screen. The design methodology is readily applicable to other field strengths or nuclei. Improvements over existing ^3^He coils were found in terms of sensitivity and transmit field homogeneity, an important feature in HP MRI. Magn Reson Med 60:431–438, 2008. © 2008 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Susceptibility effects in hyperpolarized
✍ Martin H. Deppe; Juan Parra-Robles; Salma Ajraoui; Steven R. Parnell; Matthew Cl 📂 Article 📅 2009 🏛 John Wiley and Sons 🌐 English ⚖ 680 KB

## Abstract ## Purpose To compare susceptibility effects in hyperpolarized ^3^He lung MRI at the clinically relevant field strengths of 1.5T and 3T. ## Materials and Methods Susceptibility‐related B~0~ inhomogeneity was evaluated on a macroscopic scale by B~0~ field mapping via phase difference.