Assessment of prosthetic aortic valve performance by magnetic resonance velocity imaging

โœ Scribed by Botnar, R


Publisher
Springer
Year
2000
Tongue
English
Weight
626 KB
Volume
10
Category
Article
ISSN
0968-5243

No coin nor oath required. For personal study only.

โœฆ Synopsis


Objectives: Magnetic resonance (MRI) velocity mapping was used to evaluate non-invasively the flow profiles of the ascending aorta in normal volunteers and in patients with an aortic (mechanical) valve prosthesis.

Background: In patients with artificial aortic valves the flow profile in the ascending aorta is severely altered. These changes have been associated with an increased risk of thrombus formation and mechanical hemolysis.

Methods: Velocity profiles were determined 30 mm distal to the aortic valve in six healthy volunteers and seven patients with aortic valve replacement (replacement within the last 2 years) using ECG triggered phase contrast MRI. Peak flow, mean flow and mean reverse flow were measured in intervals of 25 ms during the entire heart cycle. Systolic reverse flow, end-systolic closing and diastolic leakage volume were calculated for all subjects.

Results: Peak flow velocity during mid-systole was significantly higher in patients with valvular prosthesis than in normals (mean9SD, 1.9 90.4 m/s vs. 1.29 0.03 m/s, PB 0.001) with a double peak and a zone of reversed flow close to the inner (left lateral) wall of the ascending aorta of the patients. Closing volume was significantly larger in patients than in controls ( -3.3 91.2 ml/beat vs. -0.990.5 ml/beat; PB0.001). There was reverse flow during systole in valvular patients amounting to 15.7 9 6.7% of total cardiac output compared to 2.3 9 1.2% in controls (PB 0.001). Diastolic mean flow was negative in patients after valve replacement but not in controls ( -11.09 15.2 ml/beat vs. 6.8 93.2 ml/beat; P B 0.01).

Conclusions:

The following three major quantitative observations have been made in the present study: (1) Mechanical valve prostheses have an increased peak flow velocity with a systolic reverse flow at the inner (left lateral) wall of the ascending aorta.

(2) A double peak flow velocity pattern can be observed in patients with bileaflet (mechanical) prosthesis. (3) The blood volume required for leaflet closure and the diastolic leakage blood volume are significantly higher for the examined bileaflet valve than for native heart valves.


๐Ÿ“œ SIMILAR VOLUMES


Diagnosis and assessment of mitral and a
โœ Leslie Mitchell; Jeremy P. R. Jenkins; Yvonne Watson; Derek J. Rowlands; Ian Ish ๐Ÿ“‚ Article ๐Ÿ“… 1989 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 920 KB

Seventy-six aortic and mitral valves, in 44 patients and 5 normal volunteers, were studied by Cine-Flow MRI (on a0.26-T superconducting magnet system), utilizingcompound oblique imaging planes and a Field Echo Even Rephasing sequence. All patients had had cardiac catheterization and echocardiography

Improved aortic pulse wave velocity asse
โœ Jos J.M. Westenberg; Albert de Roos; Heynric B. Grotenhuis; Paul Steendijk; Denn ๐Ÿ“‚ Article ๐Ÿ“… 2010 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 367 KB

Purpose: To evaluate the accuracy and reproducibility of aortic pulse wave velocity (PWV) assessment by in-plane velocity-encoded magnetic resonance imaging (MRI). Materials and Methods: In 14 patients selected for cardiac catheterization on suspicion of coronary artery disease and 15 healthy volun

Left atrial appendage flow velocities: a
โœ Kai Muellerleile; Arian Sultan; Michael Groth; Daniel Steven; Imke Drewitz; Bori ๐Ÿ“‚ Article ๐Ÿ“… 2011 ๐Ÿ› BioMed Central ๐ŸŒ English โš– 232 KB

To evaluate the use of velocity encoded (VENC) magnetic resonance imaging (MRI) for measurements of left atrial appendage (LAA) flow velocities in comparison with transesophageal echocardiography (TEE).