## Abstract The accuracy of MR phase contrast (PC) velocity measurements for unsteady flow has been quantitatively assessed. Spatially resolved velocity fields were measured in a long straight tube using a gated PC technique. and the resulting MR PC velocity data were compared with velocities deriv
Assessing the accuracy of two-dimensional phase-contrast MRI measurements of complex unsteady flows
✍ Scribed by Yannis Papaharilaou; Denis J. Doorly; Spencer J. Sherwin
- Publisher
- John Wiley and Sons
- Year
- 2001
- Tongue
- English
- Weight
- 318 KB
- Volume
- 14
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Two‐dimensional phase‐contrast MRI measurements of complex unsteady flows have been assessed for accuracy, together with procedures used to improve the precision of the measurements. Velocity measurements of single harmonic sinusoidal flow in a rigid bypass graft model with a fully three‐dimensional geometry were compared to an accurate numerical solution of the Navier‐Stokes equations for the same flow. Axial velocity profiles from the MRI were compared with the computational data, and instantaneous root mean square (rms) differences were calculated. Despite the complexity of the flow, with the aid of phase angle dynamic range extension, a spatially and temporally averaged rms error of between 7.8% and 11.5%, with respect to the spatially and temporally averaged velocity, was achieved. Spin saturation primarily and phase dispersion secondarily in complex transient recirculation zones were found to be significant contributors to overall error. Cross flow effects were also investigated but were of lesser significance. The result confirms the suitability of the technique for measuring complex unsteady flows. J. Magn. Reson. Imaging 2001;14:714–723. © 2001 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract In this work the effects of noise, resolution, and velocity (flow) on the measurement of intravascular pressure from phase‐contrast (PC) MRI are discussed. To elucidate these effects, we employed an axisymmetric geometry that enabled us to calculate pressures in <2 min on a Sun Ultra SP
## Abstract The accuracy of volume flow rate measurements obtained with phase‐contrast methods was assessed by means of computer simulation and in vitro experiments. Factors studied include (a) the partial‐volume effect due to voxel dimensions relative to vessel dimensions and orientation and (b) i
## Abstract Phase‐contrast (PC) magnetic resonance imaging (MRI) flow measurements suffer from the effect of the point spread function (PSF) due to the limited sampling of __k__‐space. The PSF, which in this case is a __sinc__ function, deforms the flow profile and forms a ringing pattern around th