## Abstract The application of __T__~1__P__~ as an alternative contrast parameter in high‐field magnetic resonance histology (MRH) has been investigated. Spectroscopic measurements of __T__~1__P__~ were performed on 5.75% agar and 1.0 mM MnCl~2~ phantoms at 9.4 T to validate the accuracy of the ima
Application of the keyhole technique to T1ρ relaxation mapping
✍ Scribed by Andrew J. Wheaton; Arijitt Borthakur; Ravinder Reddy
- Publisher
- John Wiley and Sons
- Year
- 2003
- Tongue
- English
- Weight
- 200 KB
- Volume
- 18
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Purpose:
To demonstrate the feasibility of using the keyhole technique to minimize error in a least squares regression estimation of t(1rho) from magnetic resonance (mr) image data.
Materials and methods:
The keyhole method of partial k-space acquisition was simulated using data from a virtual phantom and mr images of ex vivo bovine and in vivo human cartilage. t(1rho) maps were reconstructed from partial k-space (keyhole) image data using linear regression, and error was measured with relation to t(1rho) maps created from the full k-space images. an error model was created based on statistical theory and fitted to the error measurements.
Results:
T(1rho) maps created from keyhole images of a human knee produced levels of error on the order of 1% while reducing standard image acquisition time approximately by half. the resultant errors were strongly correlated with expectations derived from statistical theory.
Conclusion:
The error model can be used to analytically optimize the keyhole method in order to minimize the overall error in the estimation of the relaxation parameter of interest. the keyhole method can be generalized to significantly expedite all forms of relaxation mapping.
📜 SIMILAR VOLUMES
A method for the volume-and resonance line-selective determination of the longitudinal relaxation time in the rotating frame, T,,, is described. The spin-lock pulse intrinsic to the T,, sequence simultaneously replaces the first slice-selective pulse of the VOSY method for localized spectroscopy. Th
## Abstract ## Purpose: To determine the relationship between changes in the extracellular matrix (ECM) and T~1ρ~ and T~2~ values in vivo. The ECM is composed of proteoglycan (PG), collagen, and water. It has been unclear which of the ECM constituents affects T~1ρ~ and T~2~ mapping in living human
## Abstract Three‐dimensional spin‐lattice relaxation time in the rotating frame (3D‐T~1ρ~) with parallel imaging at 3.0T was implemented on a whole‐body clinical scanner. A 3D gradient‐echo sequence with a self‐compensating spin‐lock pulse cluster was combined with generalized autocalibrating part
Solid-state spin-lattice relaxation in the rotating frame permits the investigation of dynamic processes with correlation times in the range of microseconds. The relaxation process in organic solids is driven by the fluctuation of the local magnetic field due to the dipole-dipole interaction of the