Analytical measurement of two nuclides using a liquid scintillation counter
β Scribed by Haruo Fujii; Makoto Takiue; Hiroaki Ishikawa
- Publisher
- Elsevier Science
- Year
- 1989
- Weight
- 140 KB
- Volume
- 40
- Category
- Article
- ISSN
- 0883-2889
No coin nor oath required. For personal study only.
π SIMILAR VOLUMES
It was recently demonstrated that in a liquid scintillation spectrometer hard p-emitters (""P, 24Na, lo6Rh, 144Pr, Q"Y, 40K, \*04T1, s"Sr, 13'Cs, 42K) can be measured in aqueous solutions without the addition of scintillator (l-3). This is due to the fact that fast fl-particles in water give rise to
Cerenkov counting efficiency varies with colour quenching and sample turbidity. The activity in a plastic vial can be determined accurately with a colour quenching correction technique, regardless of the presence or absence of turbidity in a sample. On the other hand, the error of the measured activ