Analysis of trichothecene mycotoxins by mass spectrometry and tandem mass spectrometry
β Scribed by Ronald D. Plattner; Marian N. Beremand; Richard G. Powell
- Publisher
- Elsevier Science
- Year
- 1989
- Tongue
- French
- Weight
- 954 KB
- Volume
- 45
- Category
- Article
- ISSN
- 0040-4020
No coin nor oath required. For personal study only.
β¦ Synopsis
To illustrate tandem mass spectrometry (MS/MS) as an alternative to GC/MS for the analysis of trichothecenes in solid and liquid cultures, diacetoxyscerpenol (DAS) levels were measured by GC/MS and by three MS/MS protocols. GC/MS had the lowest detection limit and better precision than the MS/MS procedures; however, GC/MS required 20-30 minutes per analysis. MS/MS was more rapid, requiring less than two minutes per analysis. DAS could be measured accurately at low nanogram levels directly from filtrates of liquid cultures. Extraction of culture filtrates simplified the identification by effectively removing an interfering matrix component which had an ion at the same m/z as the protonated DAS molecule. Use of the GC as an inlet for MS/MS was an effective way to minimize physical manipulations necessary for sample introduction. All three MS/MS procedures had similar detection limits and precisions. Occurrence levels of three trichothecenes, T-2 toxin, neosolaniol, and DAS were measured in a single analysis by sequential selection of parents for each trichothecene.
π SIMILAR VOLUMES
Tetrodotoxin (TTX), a powerful sodium channel blocker, usually exists as a mixture of its analogs (TTXs) in natural sources. Due to the structural variation, some analogs are difficult to detect using the postcolumn liquid chromatography-fluorescent detection (LC-FLD) system. Liquid chromatography/e
The use of lowenergy (7 eV Elnb) reactive collisions with ammonia provides a highly specific method for the characterization of trichothecenes since the collisions between the protonated molecule and ammonia not only produce ions formed by collisionally activated dissociation but ammonium adduct ion
Low-energy negative-ion electrospray mass spectrometry (ESI-MS) and ESI-MS/MS were used to characterize saturated and unsaturated fatty acids. The carbon number and degree of unsaturation of fatty acids were determined using ESI-MS, and MS/MS was used to localize some double bond positions of mono-a