## Abstract A novel LC–fluorescence detection method based on the use of a monolithic column for the determination of norfloxacin, ciprofloxacin, and enrofloxacin antibiotic residues in environmental waters was developed. Fluoroquinolones (FQs) were isocratically eluted using a mobile phase consist
Analysis of microcystins by capillary high performance liquid chromatography using a polymethacrylate-based monolithic column
✍ Scribed by Congying Gu; Li Lin; Xiaodong Chen; Jinping Jia; Dan Wu; Nenghu Fang
- Publisher
- John Wiley and Sons
- Year
- 2007
- Tongue
- English
- Weight
- 945 KB
- Volume
- 30
- Category
- Article
- ISSN
- 1615-9306
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
A polymethacrylate‐based monolithic column was prepared and its application to the separation of three kinds of similar microcystins (MCs) in capillary high performance liquid chromatography (capillary‐HPLC) with ultraviolet detection was studied. The monolithic matrix contains both hydrophobic and cation‐exchange interaction sites. Factors influencing the separation performance have been investigated. A baseline separation could be achieved by means of Tris(hydroxymethyl)aminomethane buffer of mildly alkaline pH and acetonitrile as the mobile phases in less than 5 min. The calibration curves were linear with a correlation coefficient r >0.9990 over a range of 0.25–18.00 mg/L. This method was successfully applied to the separation of microcystins from other compounds in spiked uncontaminated lake water after performing solid‐phase extraction. The whole procedure provided low LODs, e. g. the LODs for MC‐LR, MC‐YR, and MC‐RR were found to be 0.49, 0.67, 0.30 μg/L, respectively. The LODs, precision, efficiency, and the results obtained for the real samples demonstrate the potential of polymethacrylate‐based monolithic columns as fast separation tools for routine use in the monitoring of microcystins in real water samples.
📜 SIMILAR VOLUMES
Porous graphitic carbon (PCC) is a strong reversed-phase absorbent with retention characteristics similar to those of alkyl bonded silicas, e.g. octadecylsilane (ODS)im4. However, unlike silica gel it offers a number of critically important advantages. These include the absence of residual silanol g
## Abstract A Chromolith Performance octadecylsilyl (ODS) monolithic silica column (Merck) was compared with a conventional microparticulate ODS‐bonded silica column in the high‐performance liquid chromatography separation of natural polyprenols. A system comprising two connected monolithic columns
Rapid high-performance liquid chromatography (HPLC) of polystyrenes, poly(methyl methacrylates), poly(vinyl acetates), and polybutadienes using a monolithic 50 ϫ 4.6 mm i.d. poly(styrene-co-divinylbenzene) column have been carried out. The separation process involves precipitation of the macromolecu
## Abstract In this work, the simultaneous analysis of five triazolopyrimidine sulfoanilide herbicides (flumetsulam, florasulam, metosulam, cloransulam‐methyl, and diclosulam) by HPLC using UV detection and a C~18~ monolithic column is proposed. The mobile phase which was composed of ACN, water, an