𝔖 Bobbio Scriptorium
✦   LIBER   ✦

An evaluation of the robustness of the Tanaka characterization protocol for reversed-phase liquid chromatography columns

✍ Scribed by Patrik Petersson; Melvin R. Euerby


Publisher
John Wiley and Sons
Year
2005
Tongue
English
Weight
362 KB
Volume
28
Category
Article
ISSN
1615-9306

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

The robustness of the popular ‘Tanaka’ LC column characterization protocol has been evaluated by the statistical tools of reduced factorial design, multiple linear regression and principal component analysis. These have shown that in order to obtain reliable and reproducible results, it is especially important to control the methanol content, the temperature and, in the case of the total ion‐exchange capacity test (α~B/P~ pH 7.6), the pH of the mobile phase. In particular, the hydrophobicity tests (k~PB~ and α) are sensitive to small changes in methanol content. Provided that the operating parameters for the Tanaka column characterization protocol are controlled within the following experimental limits, i. e. methanol content ±0.5% v/v, temperature ±3°C, pH ±0.10 and buffer concentration ±2.0 mM, it is feasible to distinguish between RP materials that possess selectivity differences larger than their batch‐to‐batch reproducibility. These experimental requirements can be easily met by current LC instrumentation. Hence, the validity of the Tanaka testing protocol for characterizing columns has been verified.


📜 SIMILAR VOLUMES


A trapping column for the coupling of re
✍ Vreuls, Jolan J. ;Goudriaan, Vincent P. ;Th, Udo A. ;de Jong, Gerhardus J. 📂 Article 📅 1991 🏛 John Wiley and Sons 🌐 English ⚖ 611 KB

## Abstract A new technique for coupling reversed‐phase liquid chromatography (RPLC) with gas chromatography is described. A fraction eluting from an RPLC column is trapped on a short column packed with polymeric adsorbent. After the mobile phase has been displaced with water, the analytes are deso

Comparison of the performance of differe
✍ M. Cledera-Castro; A. Santos-Montes; R. Izquierdo-Hornillos; R. Gonzalo-Lumbrera 📂 Article 📅 2007 🏛 John Wiley and Sons 🌐 English ⚖ 595 KB

## Abstract A systematic optimization of the HPLC separation of a mixture containing 11 pollutant phenols (PPs) using a Hypersil ODS (250 mm×4.6 mm, 5 μm) column and UV‐DAD detection has been carried out. The binary mobile phases used were obtained by mixing 50 mM phosphate (pH = 3.0) and methanol,

Determination of the purity of phenoxyme
✍ Milada Doležalová; Hana Čápová; Roman Jobánek 📂 Article 📅 2003 🏛 John Wiley and Sons 🌐 English ⚖ 370 KB

## Abstract A micellar electrokinetic chromatographic and a fast reversed‐phase liquid chromatographic method have been developed for determination of the purity of phenoxymethylpenicillin. The optimized running buffer composition was 40 mM phosphate–borate–125 mM SDS–3.5% (__v/v__) methanol. The H