𝔖 Bobbio Scriptorium
✦   LIBER   ✦

An Er:YAG laser endoscopic fiber delivery system for lithotripsy of salivary stones

✍ Scribed by Joshua Raif; Michael Vardi; Oded Nahlieli; Israel Gannot


Publisher
John Wiley and Sons
Year
2006
Tongue
English
Weight
227 KB
Volume
38
Category
Article
ISSN
0196-8092

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Background and Objectives

Endoscopic applications of Erbium:YAG lasers are still very limited due to lack of appropriate fiber delivery capabilities. Recent reports on potential advantages of this laser for lithotripsy of ureteral stones prompted us to develop an Er:YAG fiber delivery system for endoscopic lithotripsy of salivary stones. We report on the development of this system and its clinical use on 17 patients.

Study Design/Materials and Methods

Ho:YAG and Er:YAG laser fragmentation performances were initially compared. Optimal laser parameters for lithotripsy of salivary stones were then established ex vivo using a commercial dental Er:YAG laser (Lumenis Opusdent 20). Metal hollow waveguides optimized for Er:YAG laser transmission were end sealed with a polished sapphire rod of 0.63 mm diameter and designed to adapt to the Opusdent laser and to a Storz sialoendoscope. The system was tested ex vivo for durability and clinical compatibility at input energies up to 700 mJ, 10–20 Hz. Following Helsinki approval the system was clinically tested on 17 patients with sialolithiasis.

Results

Lithotripsy threshold was around 80 mJ/pulse (26 J/cm^2^) while efficient fragmentation, with microscopic fragments, was observed at an output energy range of 150–300 mJ/pulse. At 10 Hz, fragmentation rates of about 1.8 mm^3^/second were achieved enabling lithotripsy of a 6 mm stone in about 2 minutes. Front surface damage to the sapphire rod occurred but did not contribute to significant loss in fragmentation efficiency. Of the 21 stones treated clinically, 5 were fully fragmented, 7 were prepared for extraction by mini forceps, and 9 were released from surrounding soft tissues for subsequent removal. Fifteen of the 18 treated glands returned to normal function without any symptoms.

Conclusions

The Er:YAG endoscopic delivery system described is a clinically viable and cost‐effective device for a range of hard and soft tissue wet field applications accessible through rigid or semi‐rigid endoscopes. Further improvements in the waveguide may allow access also through fully flexible endoscopes. Lasers Surg. Med. 38:580–587, 2006. © 2006 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Comparison of fluoride and sapphire opti
✍ Jinze Qiu; Joel Teichman; Tianyi Wang; Badr Elmaanaoui; David Gamez; Thomas E. M 📂 Article 📅 2010 🏛 John Wiley and Sons 🌐 English ⚖ 219 KB

## Abstract The long‐pulse (200–350 μs) Holmium: YAG (Ho: YAG) laser (λ = 2.12 μm) is used extensively in urology for laser lithotripsy. The long‐pulse Erbium: YAG (Er: YAG) laser (λ = 2.94 μm) fragments urinary calculi up to 5 times more efficiently than the Ho: YAG laser, however, no optical fibe

Hybrid germanium/silica optical fibers f
✍ Charles A. Chaney; Yubing Yang; Nathaniel M. Fried 📂 Article 📅 2004 🏛 John Wiley and Sons 🌐 English ⚖ 197 KB

## Abstract ## Background and Objectives Endoscopic applications of the erbium (Er):YAG laser have been limited due to the lack of an optical fiber delivery system that is robust, flexible, and biocompatible. This study reports the testing of a hybrid germanium/silica fiber capable of delivering E