An apparatus for measuring the mössbauer effect under pressure at low temperatures
✍ Scribed by G.N. Stepanov; V.N. Panyushkin
- Publisher
- Elsevier Science
- Year
- 1977
- Tongue
- English
- Weight
- 204 KB
- Volume
- 17
- Category
- Article
- ISSN
- 0011-2275
No coin nor oath required. For personal study only.
✦ Synopsis
The existence of dynamic temperature control, the good thermal contact between heater and cryostat, the large power dissipated in the heater transistor enable a given temperature to be reached rapidly and for it to be maintained accurately. A temperature stability in the cryostat of not worse than 0.01 K can be achieved with the system, with a time of 5 s for establishing it when the temperature is changed by 10 K and the mass of the copper specimen holder is ~ 30 g.
Operating P215 Ge transistors as heating elements for low temperature cryostats over two years has shown their high reliability for many temperature cycles in the range 77 to 340 K.
References
📜 SIMILAR VOLUMES
An apparatus has been developed for the measurement of equilibrium vapour and liquid composition and refractive index data at low temperature and high pressure. Equilibrium is achieved with an externally coupled magnetic mixing device. Phase samples are removed for analysis through specially designe
would need to be trapped in order to bring the viscosity results back to the correct value. A clue to this can be found from observations of the specific heat 4 and adsorption characteristics 5 of He 3-tte 4 mixture films, from which it was inferred that the He 3 component of such films resides main
Comparative ebulliometry has been used to determine the vapour pressures of Ž . Ž perfluoromethylcyclopentane between T s 260 K, p s 10.6 kPa and T s 434 K, p s . 1590 kPa . The measurements at low temperatures were made using a new stainless steel ebulliometer with a condenser that automatically ad
This paper describes the performance characteristics of a new test apparatus and the measured results using the apparatus. It has been developed on the basis of Harmathy's curve-fitting method to measure, in a short time, the physical properties of the nonmetallic materials for a cryogenic storage t