The high resolution of capillary zone electrophoresis/mass spectrometry (CZE/MS) offers a promising technique to characterize biomolecules in pharmaceutical and biotechnology industries. A novel capillary zone electrophoresis/electrospray ionization time-of-flight mass spectrometry (CZE/ESI-TOF MS)
A sonic spray interface for capillary electrophoresis/mass spectrometry
β Scribed by Yukiko Hirabayashi; Atsumu Hirabayashi; Hideaki Koizumi
- Publisher
- John Wiley and Sons
- Year
- 1999
- Tongue
- English
- Weight
- 70 KB
- Volume
- 13
- Category
- Article
- ISSN
- 0951-4198
No coin nor oath required. For personal study only.
β¦ Synopsis
We have developed a sonic spray ionization interface that enables coupling of capillary electrophoresis (CE) and mass spectrometry (MS). In sonic spray ionization, a sample solution is sprayed at any solution-flow rate from a sample-introduction capillary with a high-speed gas flow coaxial to the capillary, and ions are formed at atmospheric pressure. Therefore, it is used with a wide range of buffer solutions regardless of the conductivity of the solutions. However, the pressure around the tip of the sample-introduction capillary is reduced by the high-speed gas flow, so the solution is pumped into the capillary at a flow rate above 0.1 mL/ min due to the difference of pressure between the two ends of the capillary. Since the solution-flow rate in CE is much lower than this pumping rate, the resolution of CE separation is expected to be decreased by the pumping effect when an electrophoresis capillary is connected directly to the sample-introduction capillary. To avoid this in our CE/MS interface, we added a buffer reservoir between the sample-introduction capillary of the ion source and the electrophoresis capillary. We confirmed that this prevents the solution in the electrophoresis capillary being pumped by the pressure difference. Furthermore, we have demonstrated analysis by CE/MS with a mobile-phase buffer containing 15 mM of phosphate by filling the buffer reservoir with an acetic-acid solution as a substitute for the mobile-phase buffer. This increased the ion intensity 100fold by enhancing the evaporation of charged droplets produced by the spray.
π SIMILAR VOLUMES
## Modified liquid junction interface for nonaqueous capillary electrophoresis-mass spectrometry Electrospray ionization (ESI) is the most widely used ionization method in on-line coupling of capillary electrophoresis-mass spectrometry (CE-MS). The conventional coaxial sheath flow electrospray int
The interfacing of capillary electrophoresis (CE) with mass spectrometry (MS) is well established and may be accomplished by use of either a coaxial arrangement or by employing a liquid T-junction. In both these interfaces a make-up flow is introduced. This is required because of the mismatch in flo
Explosive vaporization and mist formation occurs when an aqueous solution effusing from the tip of the stainless steel capillary (inner diameter: 0.1 mm) was irradiated from the opposite side of the capillary by a 10.6 mm infrared laser focused to %0.1 mm. Application of a high voltage (3-4 kV) to t
The construction of a sheathless interface for capillary electrophoresis-electrospray ionization mass spectrometry (CE-ESI-MS), for operation with a Z-Spray source on a Micromass Quattro-LC triple quadrupole mass spectrometer is described. Designing the interface involved machining a probe compatibl