Knowledge of the spatial distribution of transmission field B(1)(+) and reception sensitivity maps is important in high-field (≥3 T) human magnetic resonance (MR) imaging for several reasons: these include post-acquisition correction of intensity inhomogeneities, which may affect the quality of imag
A phase-sensitive method of flip angle mapping
✍ Scribed by Glen R. Morrell
- Publisher
- John Wiley and Sons
- Year
- 2008
- Tongue
- English
- Weight
- 548 KB
- Volume
- 60
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
A radiofrequency (RF) excitation scheme is presented in which flip angle is encoded in the phase of the resulting excitation. This excitation is implemented with nonselective hard pulses, and is used to give flip angle maps over three‐dimensional volumes. This phase‐sensitive B1 mapping excitation can be combined with various acquisition methods such as gradient recalled echo (GRE) and echo‐planar (EP) readouts. Imaging time depends primarily on the readout method, and is roughly equivalent to the imaging time of conventional double‐angle techniques for three‐dimensional acquisition. The phase‐sensitive method allows imaging over a much wider range of flip angles than double‐angle methods. Phantom and in vivo results are presented comparing the phase‐sensitive method with the conventional double‐angle method, demonstrating the ability of the phase‐sensitive method to measure a wider range of flip angles than double‐angle methods. Magn Reson Med 60:889–894, 2008. © 2008 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract It is demonstrated that the stability of the Carr‐Purcell‐Mei‐boom‐Gill (CPMG) sequence reflects the existence of a steady state solution to the Bloch equations in the absence of __T__~2~ and __T__~1~ decay. The steady state theory is then used to evaluate the performance of low flip an
setting the correct rf power levels to achieve YO" and 180" rf pulses for MR imaging is to peak the echo amplitude of a rf spin-echo sequence. The echo amplitude of this 01-201 pulse sequence is proportional to sin3( a) and has a relatively broad maximum. Recently another method for setting the rf f
## Abstract A new method has been developed for fast image‐based measurements of the transmitted radiofrequency (RF) field. The method employs an actual flip‐angle imaging (AFI) pulse sequence that consists of two identical RF pulses followed by two delays of different duration (TR~1~ < TR~2~). Aft