A new configuration for coupling purge-and-trap/cryogenic focusing/capillary column gas chromatography with splitless injection mode

โœ Scribed by Liu, Shili ;Kang, Jiangshi ;Strother, Dennis L. ;Carley, Robert J. ;Stuart, James D.


Publisher
John Wiley and Sons
Year
1989
Tongue
English
Weight
470 KB
Volume
12
Category
Article
ISSN
0935-6304

No coin nor oath required. For personal study only.

โœฆ Synopsis


A new configuration for coupling a purge-and-trap unit to a capillary column gas chromatograph via a cryogenic focusing interface has been developed. In this configuration, the precolumn of the cryogenic focusing interface was inserted through the septum of a split/splitless injection port where it served as both sample transfer and carrier gas supply lines. The injection port of the gas chromatograph was modified by plugging the carrier gas and the septum purge lines. This configuration allowed for the desorption of analytes at high flow rates while maintaining low, analytical-column flow rates which are necessary for optimum capillary column operation. The capillary column flow rate is still controlled by the column backpressure regulator. Chromatograms of purgeable aromatics exhibited improved resolution, especially for early eluting components compared to those obtained by direct liquid injection using the normal splitless injection mode. Quantitative sample transfer to the analytical column afforded excellent linearity and reproducibility of compounds studied.


๐Ÿ“œ SIMILAR VOLUMES


Analysis of headspace samples using off-
โœ Haynes, L. V. ;Steimle, A. R. ๐Ÿ“‚ Article ๐Ÿ“… 1987 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 421 KB ๐Ÿ‘ 1 views

Gas chromatographic equipment and procedures are described for automated splitless injection of pseudo-static headspace samples collected externally onto a sorbent trap. The GC microprocessor controls, in sequence, carrier gas backflushing of the sorbent trap for water removal, splitless thermal des