A microcomputer-controlled system for stimulation and acquisition of evoked potentials

✍ Scribed by Raziel Haimi-Cohen; Arnon Cohen


Publisher
Elsevier Science
Year
1984
Tongue
English
Weight
710 KB
Volume
17
Category
Article
ISSN
0010-4809

No coin nor oath required. For personal study only.

✦ Synopsis


A system, controlled by a microcomputer (LSI-1 l/23 with appropriate peripherals) for conducting evoked potential (EP) experiments, is described. In each experimental trial, a stimulus is applied and multichannel EEG is sampled, displayed, and stored in an output file. Any required length of pre-and poststimulus sampling periods is available. Stimuli are triggered by a human operator, aided by an on-line graphic display of sampled data. Automatic artifact monitoring is used for rejecting bad trials. Various stimulators may be used (simultaneously) and stimulus features (duration, power, number of pulses) may be changed in each trial. Lists of trial parameters, such as stimulus features and subject's response, are incorporated in the output file to facilitate cross-section analysis. The system was designed for the needs of pain EP research, but its flexibility and modularity make it useful in other areas of EP research.


πŸ“œ SIMILAR VOLUMES


A microcomputer control system for the m
✍ K.R. Abbott; P. Doig πŸ“‚ Article πŸ“… 1986 πŸ› Elsevier Science 🌐 English βš– 899 KB

A microcomputer-based system for the measurement of fatigue crack growth dnldn versus cyclic stress intensity factor AK data using compact-tension test specimens is described. The procedure has been developed to allow automatic measurement of crack growth rate under any specified combination and seq

Laboratory instrument interface system (
✍ Douglas R. Danforth; Richard L. Stouffer; Michael Bosnos πŸ“‚ Article πŸ“… 1985 πŸ› Elsevier Science 🌐 English βš– 775 KB

A simple method of interfacing clinical and research laboratory equipment with a microcomputer is described. A four-channel buffer system has been constructed which stores data generated from laboratory instruments and then transmits the data directly to a microcomputer. The system is highly flexibl