Quantitative assessment of Magnetic Resonance Imaging (MRI) lesion load of patients with multiple sclerosis (MS) is the most objective approach for a better understanding of the history of the pathology, either natural or modi®ed by therapies. To achieve an accurate and reproducible quanti®cation of
A method of radio-frequency inhomogeneity correction for brain tissue segmentation in MRI
✍ Scribed by L.Q. Zhou; Y.M. Zhu; C. Bergot; A.-M. Laval-Jeantet; V. Bousson; J.-D. Laredo; M. Laval-Jeantet
- Publisher
- Elsevier Science
- Year
- 2001
- Tongue
- English
- Weight
- 433 KB
- Volume
- 25
- Category
- Article
- ISSN
- 0895-6111
No coin nor oath required. For personal study only.
✦ Synopsis
An automatic method of correcting radio-frequency (RF) inhomogeneity in magnetic resonance images is presented. The method considers that image intensity variation due to radio-frequency inhomogeneity contains not only low frequency components, but also high frequency components. The variation is regarded as a multiplication of low frequency (capacity variation of coil) and the frequency of object (true image). The ef®ciency of the proposed method is illustrated with the aid of both phantom and physical images. The impact of the inhomogeneity correction on brain tissue segmentation is studied in detail. The results show signi®cant improvement of the tissue segmentation after inhomogeneity correction.
📜 SIMILAR VOLUMES
## Abstract A novel radio‐frequency (RF) pulse design algorithm is presented that generates fast slice‐selective excitation pulses that mitigate __B__ inhomogeneity present in the human brain at high field. The method is provided an estimate of the __B__ field in an axial slice of the brain and the
## Abstract A simple method is described that corrects for the intensity inhomogeneity experienced when using endorectal surface coils in prostate imaging. The method uses a series of proton‐density‐weighted images, which are acquired to demonstrate the sensitivity profile of the coil and used to c