✦ Synopsis


Abstract

Purpose

To improve the performance of 7T head coils over the rostral head regions. Due to radiofrequency (RF) field/tissue interactions, the RF magnetic field profile produced by 7T volume head coils is very inhomogeneous, with enhanced sensitivity near the center of the human brain and substantially reduced in the periphery.

Materials and Methods

Two head‐sized quadrature volume coils of similar diameters but substantially different lengths (17 and 10 cm) were constructed and tested using a 7T Varian Inova system.

Results

Experimental data demonstrated that by using a shorter volume head‐sized coil or simply by partially moving a head out of the coil, coil efficiency near the top of a head can be improved by 20%. The homogeneity also improved, largely resulting from an increase in peripheral B~1~ values. This resulted in 10%–20% variation in axial slices located near the top of a head.

Conclusion

We have demonstrated a less deeply positioned head or substantially shorter volume coil can significantly improve coil performance and homogeneity for the rostral head at ultrahigh magnetic fields (7T and above). For studies that target superior brain regions, this coil arrangement can be highly effective. J. Magn. Reson. Imaging 2009;29:461–465. © 2009 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Eight-channel phased array coil and detu
✍ G. C. Wiggins; A. Potthast; C. Triantafyllou; C. J. Wiggins; L. L. Wald 📂 Article 📅 2005 🏛 John Wiley and Sons 🌐 English ⚖ 478 KB

## Abstract An eight‐channel receive‐only brain coil and table‐top detunable volume transmit coil were developed and tested at 7 T for human imaging. Optimization of this device required attention to sources of interaction between the array elements, between the transmit and receive coils and minim

Comparison of a 28-channel receive array
✍ Gregory Chang; Graham C. Wiggins; Ding Xia; Riccardo Lattanzi; Guillaume Madelin 📂 Article 📅 2011 🏛 John Wiley and Sons 🌐 English ⚖ 835 KB

## Abstract ## Purpose: To compare a new birdcage‐transmit, 28‐channel receive array (28‐Ch) coil and a quadrature volume coil for 7T morphologic MRI and T2 mapping of knee cartilage. ## Materials and Methods: The right knees of 10 healthy subjects were imaged on a 7T whole body magnetic resonan