Eight-channel phased array coil and detunable TEM volume coil for 7 T brain imaging
✍ Scribed by G. C. Wiggins; A. Potthast; C. Triantafyllou; C. J. Wiggins; L. L. Wald
- Book ID
- 102533648
- Publisher
- John Wiley and Sons
- Year
- 2005
- Tongue
- English
- Weight
- 478 KB
- Volume
- 54
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
An eight‐channel receive‐only brain coil and table‐top detunable volume transmit coil were developed and tested at 7 T for human imaging. Optimization of this device required attention to sources of interaction between the array elements, between the transmit and receive coils and minimization of common mode currents on the coaxial cables. Circular receive coils (85 mm dia.) were designed on a flexible former to fit tightly around the head and within a 270‐mm diameter TEM transmit volume coil. In the near cortex, the array provided a fivefold increase in SNR compared to a TEM transmit‐receive coil, a gain larger than that seen in comparable coils at 3 T. The higher SNR gain is likely due to strong dielectric effects, which cause the volume coil to perform poorly in the cortex compared to centrally. The sensitivity and coverage of the array is demonstrated with high‐resolution images of the brain cortex. Magn Reson Med 54:235–240, 2005. © 2005 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract ## Purpose: To compare a new birdcage‐transmit, 28‐channel receive array (28‐Ch) coil and a quadrature volume coil for 7T morphologic MRI and T2 mapping of knee cartilage. ## Materials and Methods: The right knees of 10 healthy subjects were imaged on a 7T whole body magnetic resonan
## Abstract ## Purpose To investigate changes in image and dynamic signal‐to‐noise ratios (SNRs) of the ΔR2\* curve, as well as magnetic susceptibility‐induced artifacts between a standard quadrature head coil and an eight‐channel phased‐array coil with and without sensitivity‐encoding (SENSE) at
## Abstract ## Purpose To evaluate the ability of a custom‐built coil design to provide improved signal‐to‐noise ratio (SNR) and less signal drop with increasing depth at the carotid artery. ## Materials and Methods Phased‐array surface coils can provide a high SNR to image the carotid vessel wa