Xanthine Biosensor Based on Didodecyldimethylammonium Bromide Modified Pyrolytic Graphite Electrode
β Scribed by Ji-Lin Tang; Xiao-Jun Han; Wei-Min Huang; Er-Kuq Wang
- Publisher
- John Wiley and Sons
- Year
- 2010
- Tongue
- English
- Weight
- 283 KB
- Volume
- 20
- Category
- Article
- ISSN
- 0256-7660
No coin nor oath required. For personal study only.
β¦ Synopsis
Abstract
The vesicle of didodecyldimethylammonium bromide (DDAB) which contained tetrathiafulvalene (TTF) was mixed with xanthine oxidase, and the mixture was cast on the pyrolytic graphite electrode. The lipid films were used to supply a biological environment resembling biomembrane on the surface of the electrode. TTF was used as a mediator because of its high electronβtransfer efficiency. A novel xanthine biosensor based on cast DDAB film was developed. The effects of pH and operating potential were explored for optimum analytical performance by using the amperometric method. The response time of the biosensor was less than 10 s. The detection limit of the biosensor was 3.2 Γ 10^β7^ mol/L and the liner range was from 4 Γ 10^β7^ mol/L to 2.4 Γ 10^β6^ mol/L.