Sixteen European countries participated in this WHO-IAEA intercomparison for which transmission CAP (College of American Pathologists) thyroid and IAEA-WHO liver phantoms were used. A total of 257 laboratories submitted 428 image evaluation reports. Overall results showed differences in performance
World health organisation inter-laboratory comparison study in 12 countries on quality performance of nuclear medicine imaging devices
✍ Scribed by V. Volodin; G. Souchkevitch; N. Racoveanu; H. Bergmann; E. Busemann-Sokole; B. Delaloye; F. Dermentzoglou; G. Georgescu; N. Herrera; W. Jasinski; Y. Kasatkin; P. Paras; R. Mould
- Publisher
- Springer
- Year
- 1985
- Tongue
- English
- Weight
- 610 KB
- Volume
- 10-10
- Category
- Article
- ISSN
- 0340-6997
No coin nor oath required. For personal study only.
✦ Synopsis
Twelve countries participated in the WHO intercomparison for which transmission CAP (College of American Pathologists) brain and CAP liver phantoms and emission London liver phantoms were used. A total of 157 imaging devices were tested. Overall results from the phantoms revealed a wide range of targets detected. For the CAP-phantoms target detectability dropped below 50% for target sizes less than 9 mm and below 60% for target contrast less than 0.84:1. On average one false positive and six false negative results were reported using CAP-brain phantoms and one false positive and one false negative result using CAP-liver phantoms. For the London liver phantoms containing the tissue equivalent rubber abdominal simulation the target of 1 cm was never visualised. Two targets in this phantom (2 cm and 2.5 cm) were correctly identified in 34% of studies and one of these targets in 52% of studies. Equivocal and false positive results were reported in 42%. The WHO inter-comparison demonstrated the need to establish new, or to improve the existing, quality control programmes.
📜 SIMILAR VOLUMES