𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Wear analysis of retrieved ceramic-on-ceramic articulations in total hip arthroplasty: Femoral head makes contact with the rim of the socket outside of the bearing surface

✍ Scribed by Tetsuya Yamamoto; Masanobu Saito; Masaru Ueno; Takehito Hananouchi; Yusuke Tokugawa; Kazuo Yonenobu


Publisher
John Wiley and Sons
Year
2005
Tongue
English
Weight
310 KB
Volume
73B
Category
Article
ISSN
1552-4973

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

We investigated articular bearing surfaces retrieved from three patients with ceramic‐on‐ceramic total hip arthroplasties using high‐quality alumina. The duration of implantation was 3.5, 3.0, and 2.5 years, respectively. Dislocation of the hip joint after implantation did not occur in any patients. The retrieved prostheses were examined by stereomicroscopy and scanning electron microscopy (SEM). In the main load‐bearing area, no wear marks were seen in any sockets. However, in all three cases, stripe scars were observed at the rim of the alumina inlay under stereomicroscopy, despite the lack of indentation in the socket and the stem, which suggests impingement. Such scars were not observed on the load‐bearing surface. Under SEM, the scars showed excessive wear, including material excavation, that has not been observed in conventional hip simulator experiments. These findings are similar to those of experiments in which a simulator was programmed to distract the femoral head from the socket. The present results suggest that the femoral head can separate from the socket without neck‐socket impingement in vivo. © 2005 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater


📜 SIMILAR VOLUMES


The wear of a polyethylene socket articu
✍ Hamada, Yoshiki ;Horiuchi, Tadahiro ;Sano, Yayoi ;Usui, Ikuo 📂 Article 📅 1999 🏛 John Wiley and Sons 🌐 English ⚖ 515 KB

Wear of yttria-zirconia (zirconia) in the femoral head was investigated in mature mongrel dogs weighing 10 to 13 kg. Two dogs, which were used as a control group, were sacrificed 18 months after implantation of the uncemented modular hip system with an alumina ceramic (alumina) femoral head. A zirco