This paper describes the design criteria and construction of a compact Fabry-Perot cavity spectrometer with frequency modulation (FM) signal recovery, for use in the 70-220 GHz region. A solid state Gunn oscillator was frequency locked to a radiofrequency synthesiser and used as the primary source.
โฆ LIBER โฆ
Wavelength-modulated fabry-perot interferometry for quantitation of trace gas components
โ Scribed by Bernard C. Yip; Edward S. Yeung
- Book ID
- 104100516
- Publisher
- Elsevier Science
- Year
- 1985
- Tongue
- English
- Weight
- 351 KB
- Volume
- 169
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
โฆ Synopsis
The change in refractive index of a gas sample caused by molecular absorption is monitored by Fabry-Perot interferometry in a continuous flow system. To reduce the background noise level from other absorption processes, wavelength modulation is incorporated into the excitation laser. An enhancement in the signal-to-noise ratio of up to 146 s is found for the present method relative to intensity-modulated schemes or Stark-modulated schemes.
๐ SIMILAR VOLUMES
Quantitative millimetre wave spectrometr
โ
N.D. Rezgui; J. Allen; J.G. Baker; J.F. Alder
๐
Article
๐
1995
๐
Elsevier Science
๐
English
โ 744 KB