Voltammetric Determination of Nicotine at Poly(4-Amino-3-Hydroxynaphthalene Sulfonic Acid)-Modified Glassy Carbon Electrode
β Scribed by Alemnew Geto; Maereg Amare; Merid Tessema; Shimelis Admassie
- Publisher
- John Wiley and Sons
- Year
- 2012
- Tongue
- English
- Weight
- 547 KB
- Volume
- 24
- Category
- Article
- ISSN
- 1040-0397
No coin nor oath required. For personal study only.
β¦ Synopsis
Abstract
A poly(4βaminoβ3βhydroxynaphthalene sulfonic acid)βmodified glassy carbon electrode was developed for the voltammetric determination of nicotine. An oxidation signal at lower potential with higher current was obtained for nicotine at the polymerβmodified electrode compared to bare glassy carbon electrode suggesting the electrocatalytic effect of the polymer film. Under optimum conditions, square wave voltammetry of nicotine showed oxidation at +884β mV (vs. SCE) in phosphate buffer solution pHβ 7.5. The calibration curve was linear over the range 1.0 to 200.0β Β΅M (R^2^=0.9987) concentration of nicotine with a detection limit (S/N=3) of 8.66Γ10^β7^β molβL^β1^. Application of the proposed method to cigarette tobacco samples gave acceptable recovery results of 90.7β% to 107.2β%.