Voltammetric Determination of Nicotine at Poly(4-Amino-3-Hydroxynaphthalene Sulfonic Acid)-Modified Glassy Carbon Electrode

✍ Scribed by Alemnew Geto; Maereg Amare; Merid Tessema; Shimelis Admassie


Publisher
John Wiley and Sons
Year
2012
Tongue
English
Weight
547 KB
Volume
24
Category
Article
ISSN
1040-0397

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

A poly(4‐amino‐3‐hydroxynaphthalene sulfonic acid)‐modified glassy carbon electrode was developed for the voltammetric determination of nicotine. An oxidation signal at lower potential with higher current was obtained for nicotine at the polymer‐modified electrode compared to bare glassy carbon electrode suggesting the electrocatalytic effect of the polymer film. Under optimum conditions, square wave voltammetry of nicotine showed oxidation at +884β€…mV (vs. SCE) in phosphate buffer solution pHβ€…7.5. The calibration curve was linear over the range 1.0 to 200.0β€…Β΅M (R^2^=0.9987) concentration of nicotine with a detection limit (S/N=3) of 8.66Γ—10^βˆ’7^β€…mol L^βˆ’1^. Application of the proposed method to cigarette tobacco samples gave acceptable recovery results of 90.7 % to 107.2 %.