Vitamin C estimation with standard 1H spectroscopy using a clinical 3T MR system: Detectability and reliability within the human brain
✍ Scribed by Yi-Yu Shih; Martin Büchert; Hsiao-Wen Chung; Jürgen Hennig; Dominik von Elverfeldt
- Publisher
- John Wiley and Sons
- Year
- 2008
- Tongue
- English
- Weight
- 500 KB
- Volume
- 28
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose
To evaluate whether vitamin C is detectable using clinical routine MRS combined with LCModel analysis and verify the reliability of its estimated concentration.
Materials and Methods
In order to determine the reliability of the estimated ascorbic acid (Asc) concentrations, we analyzed 76 in vivo single voxel spectra (SVS) acquired on a 3T scanner (point‐resolved spectroscopy with TE = 30 ms, TR = 3000 ms, voxel size = 8 mL, without spectral editing) from different regions within the brain using LCModel. In addition to this we simulated multiple concentration levels by adding adapted Asc spectra to the in vivo data.
Results
Asc was successfully detected in 71 of 76 in vivo spectra with a comparable concentration ratio to that of myo‐inositol. Furthermore, we observed good linearity between added Asc concentrations and the LCModel estimates within the simulations having correlation coefficients larger than 0.985 at three different linewidth cases.
Conclusion
This study demonstrates the ability to detect vitamin C in the human brain under common clinical magnetic resonance spectroscopy (MRS) standards in combination with LCModel. Furthermore, it supports the need to include Asc in the standard MRS analysis. J. Magn. Reson. Imaging 2008;28:351–358. © 2008 Wiley‐Liss, Inc.