Vibration effects of lab equipment on dissolution testing with USP paddle method

โœ Scribed by Zongming Gao; Andrea Thies; William Doub


Publisher
John Wiley and Sons
Year
2010
Tongue
English
Weight
488 KB
Volume
99
Category
Article
ISSN
0022-3549

No coin nor oath required. For personal study only.

โœฆ Synopsis


Environmental vibration induced by laboratory equipment, building construction, or even by the analysts themselves is one of the more complicated factors affecting dissolution testing. It is difficult to control and/or calibrate by mechanical means or performance-based methods. In this study, dissolution apparatus vibration levels were measured in the frequency range from 10 to 270 Hz along all three axes using commercially available, single-axis accelerometers. The vibration distribution on the dissolution vessel plate was mapped, and acceleration was subsequently measured during dissolution runs involving NCDA#2 (10 mg prednisone) tablets using the paddle method. Several types of laboratory equipment were used to induce vibration during dissolution testing and vibration levels along the X-, Y-, and Z-axes of the vessel plate were measured in an attempt to establish possible correlation with dissolution results. In the frequency range studied, root mean square (RMS) acceleration values above 0.01 g, in either vertical or horizontal direction, typically affected dissolution results.