✦ LIBER ✦
Vascular entrapment of a ruptured intra-aortic balloon: A case report of successful removal without surgery
✍ Scribed by Cipriani, Luca ;Baldereschi, Giorgio ;Boncinelli, Lorenzo ;Marchionni, Niccolò
- Publisher
- John Wiley and Sons
- Year
- 1998
- Tongue
- English
- Weight
- 30 KB
- Volume
- 44
- Category
- Article
- ISSN
- 0098-6569
No coin nor oath required. For personal study only.
✦ Synopsis
Intra-aortic balloon pump entrapment is a rare complication that may necessitate major abdominal surgery that is potentially life threatening in the critically ill patients who require balloon counterpulsation. We report successful removal of a ruptured and entrapped intra-aortic balloon pump catheter after use of streptokinase solution to clear clots from the device. We suggest this procedure as a safer, nonsurgical method that may eliminate the need for abdominal surgery.