𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Validity, reliability, and feasibility of durometer measurements of scleroderma skin disease in a multicenter treatment trial

✍ Scribed by Merkel, Peter A. ;Silliman, Nancy P. ;Denton, Christopher P. ;Furst, Daniel E. ;Khanna, Dinesh ;Emery, Paul ;Hsu, Vivien M. ;Streisand, James B. ;Polisson, Richard P. ;Åkesson, Anita ;Coppock, John ;van den Hoogen, Frank ;Herrick, Ariane ;Mayes, Maureen D. ;Veale, Douglas ;Seibold, James R. ;Black, Carol M. ;Korn, Joseph H. ;, ;,


Publisher
John Wiley and Sons
Year
2008
Tongue
English
Weight
93 KB
Volume
59
Category
Article
ISSN
0004-3591

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Objective

To determine the validity, reliability, and feasibility of durometer measurements of skin hardness as an outcome measure in clinical trials of scleroderma.

Methods

Skin hardness was measured during a multicenter treatment trial for scleroderma using handheld digital durometers with a continuous scale. Skin thickness was measured by modified Rodnan skin score (MRSS). Other outcome data collected included the Scleroderma Health Assessment Questionnaire. In a reliability exercise in advance of the trial, 9 investigators examined the same 5 scleroderma patients by MRSS and durometry.

Results

Forty‐three patients with early diffuse cutaneous systemic sclerosis were studied at 11 international centers (mean age 49 years [range 24–76], median disease duration 6.4 months [range 0.3–23], and median baseline MRSS 22 [range 11–38]). The reliability of durometer measurements was excellent, with high interobserver intraclass correlation coefficients (ICCs) (0.82–0.92), and each result was greater than the corresponding skin site ICCs for MRSS (0.54–0.85). Baseline durometer scores correlated well with MRSS (r = 0.69, P < 0.0001), patient self‐assessments of skin disease (r = 0.69, P < 0.0001), and Health Assessment Questionnaire (HAQ) disability scores (r = 0.34, P = 0.03). Change in durometer scores correlated with change in MRSS (r = 0.70, P < 0.0001), change in patient self‐assessments of skin disease (r = 0.52, P = 0.003), and change in HAQ disability scores (r = 0.42, P = 0.017). The effect size was greater for durometry than for MRSS or patient self‐assessment.

Conclusion

Durometer measurements of skin hardness in patients with scleroderma are reliable, simple, accurate, demonstrate good sensitivity to change compared with traditional skin scoring, and reflect patients' self‐assessments of their disease. Durometer measurements are valid, objective, and scalable, and should be considered for use as a complementary outcome measure to skin scoring in clinical trials of scleroderma.


📜 SIMILAR VOLUMES


Functional index-2: Validity and reliabi
✍ Alexanderson, Helene ;Broman, Lisbet ;TollbÄck, Anna ;Josefson, Annika ;Lundberg 📂 Article 📅 2006 🏛 John Wiley and Sons 🌐 English ⚖ 298 KB

## Abstract ## Objective To revise the content of the Functional Index in myositis (FI) and to evaluate measurement properties of a revised FI. ## Methods Previously performed FI (n = 287) were analyzed for internal redundancy and consistency, and ceiling and floor effects. Content was evaluated

A multicenter, prospective, randomized,
✍ Rachel K. Hoyles; Ross W. Ellis; Jessica Wellsbury; Belinda Lees; Pauline Newlan 📂 Article 📅 2006 🏛 John Wiley and Sons 🌐 English ⚖ 156 KB

## Abstract ## Objective The lack of randomized controlled trials (RCTs) in pulmonary fibrosis in systemic sclerosis (SSc) has hampered an evidence‐based approach to treatment. This RCT was undertaken to investigate the effects of intravenous (IV) cyclophosphamide (CYC) followed by azathioprine (A