𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Validation of higher-throughput high-performance liquid chromatography/atmospheric pressure chemical ionization tandem mass spectrometry assays to conduct cytochrome P450s CYP2D6 and CYP3A4 enzyme inhibition studies in human liver microsomes

✍ Scribed by Inhou Chu; Leonard Favreau; Tony Soares; Chin-chung Lin; Amin A. Nomeir


Publisher
John Wiley and Sons
Year
2000
Tongue
English
Weight
166 KB
Volume
14
Category
Article
ISSN
0951-4198

No coin nor oath required. For personal study only.

✦ Synopsis


In the early stage of drug discovery, thousands of new chemical entities (NCEs) may be screened before a single drug candidate can be identified for development. In order to accelerate the drug discovery process, we have developed higher-throughput enzyme assays to evaluate the inhibition of cytochrome P450 isoforms 2D6 (CYP2D6) and 3A4 (CYP3A4) in human liver microsomes. The assays are based on high-performance liquid chromatography/tandem mass spectrometry (LC/MS/MS) techniques. The analysis time for each sample was reduced from approximately 20 minutes for the conventional HPLC assay to 30 seconds for the LC/MS/MS assay. For both LC/MS/MS assays, the linearity (r(2) > 0.99), precision (%CV < 15%) and accuracy (% bias <15%) for both inter- and intraday validations were satisfactory. Since the implementation of the LC/MS/MS assays, our sample throughput has increased by over 40-fold.