Validation of fast MR thermometry at 1.5 T with gradient-echo echo planar imaging sequences: phantom and clinical feasibility studies

✍ Scribed by Alexandru Cernicanu; Matthieu Lepetit-Coiffe; Joerg Roland; Christoph D. Becker; Sylvain Terraz


Publisher
John Wiley and Sons
Year
2008
Tongue
English
Weight
457 KB
Volume
21
Category
Article
ISSN
0952-3480

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

The purpose of this work was to validate in phantom studies and demonstrate the clinical feasibility of MR proton resonance frequency thermometry at 1.5 T with segmented gradient‐echo echo planar imaging (GRE‐EPI) sequences during liver tumour radiofrequency (RF) ablation. Classical GRE acquisitions and segmented GRE‐EPI acquisitions were performed at 1.5 T during simultaneous RF heating with an MR‐compatible RF electrode placed in an agar gel phantom. Temperature increments were calculated and compared with four optical temperature probe measurements using Bland– Altman analysis. In a preliminary clinical feasibility study, the rapid GRE‐EPI sequence (echo train length = 13) was used for MR temperature monitoring of RF ablation of liver tumours in three patient procedures. For phantom experiments, the Bland–Altman mean of differences between MR and optical probe temperature measurements was <0.4°C, and the 95% limits of agreement value was <1.4°C. For the in vivo studies, respiratory‐triggered GRE‐EPI acquisitions yielded a temperature accuracy of 1.3 ± 0.4°C (acquisition time = 0.6 s/image, spatial coverage of three slices/respiratory cycle). MR proton resonance frequency thermometry at 1.5 T yields precise and accurate measurements of temperature increment with both classical GRE and rapid GRE‐EPI sequences. Rapid GRE‐EPI sequences minimize intra‐scan motion effects and can be used for MR thermometry during RF ablation in moving organs. Copyright © 2008 John Wiley & Sons, Ltd.