Use of FIA systems for on-line dilution in multielement determination by ICP–MS

✍ Scribed by José Anchieta Gomes Neto; José Bento Borba Silva; Ivan Gonçalves Souza; Adilson José Curtius


Book ID
101266482
Publisher
John Wiley and Sons
Year
1999
Tongue
English
Weight
155 KB
Volume
11
Category
Article
ISSN
0895-7533

No coin nor oath required. For personal study only.

✦ Synopsis


A flow-injection (FI) system to match concentrations was used as an auto-diluter in multielement determination by inductively coupled plasma-mass spectrometry (ICP-MS). The flow system comprised loop-based injection or a timed valve that introduced a variable sample volume into a spray chamber through a standard Meinhard nebulizer of an ICP-MS.

Routinely analyzed samples such as water, plant, and steel were selected. The accuracy of multielement determination was checked against water standard reference material from the National Institute of Standards and Technology (1643d), plant standard reference material from the National Bureau of Standards (1572 citrus leaves), and steel standard reference material from the National Bureau of Standards (AISI 4340). The measuring system was calibrated with a multielement solution, yielding a