𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Two-Dimensional Mass-Transfer Model of a Flat-Plate Dialyzer with Ultrafiltration Operation

✍ Scribed by J.-W. Tu; C.-D. Ho


Publisher
John Wiley and Sons
Year
2010
Tongue
English
Weight
267 KB
Volume
33
Category
Article
ISSN
0930-7516

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

A two‐dimensional theoretical model of a flat‐plate dialyzer with ultrafiltration operation was developed in this study. The two‐dimensional velocity profile of the fluid and the concentration distribution in the membrane of the dialysis system with ultrafiltration operation were also derived. The Crank‐Nicolson method was used to numerically solve the two‐dimensional theoretical model. The influences of the retentate phase flow rate, dialysate phase flow rate, ultrafiltration flow rate and channel thickness ratio on the concentration distribution and the mass transfer rate were illustrated. A considerable mass‐transfer efficiency improvement was obtained by employing the flat‐plate dialyzer with ultrafiltration operation in comparison to the pure dialysis system without ultrafiltration operation. An experiment involving urea separation by the flat‐plate dialyzer with ultrafiltration operation was also performed to confirm the accuracy of the proposed mathematical model.