Two-dimensional echocardiographic evaluation of intracardiac transvenous pacemaker leads

✍ Scribed by Bernhard Meier; Joel M. Felner


Book ID
102330498
Publisher
John Wiley and Sons
Year
1982
Tongue
English
Weight
476 KB
Volume
10
Category
Article
ISSN
0091-2751

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

A two‐dimensional (2‐D) echocardiographic study was performed on 40 consecutive patients presenting for a routine check‐up of their transvenous pacemaker devices, in order to assess the visibility of the intracardiac portion of the pacemaker lead.

In 34 patients (85%) the presence of the lead could be demonstrated satisfactorily. In 4 of these 34 patients the entire intracardiac portion of the pacing wire was imaged, in 20 the distal fragment bearing the electrode(s) was seen, and in 10 an intermediate segment was visible. In 6 of the 40 patients (15%) the lead was not visible. In 3 of these 40 patients it was virtually impossible to image any cardiac structure. The less than 100% visibility of the pacing wire may be due to our randomized patient group with a high mean age of 75 years.

We conclude that during follow‐up examinations or when problems arise in patients with transvenous pacemakers, 2‐D echocardiography may be an alternative to roentgenograms, especially if simultaneous information about heart function is desired or septal perforation is suspected. However, chest roentgenograms are still more accurate and remain the technique of choice under normal circumstances.


📜 SIMILAR VOLUMES


Multiplane transesophageal echocardiogra
✍ Wolfgang Fehske; Werner Jung; Heyder Omran; Matthias Manz; Rainer Moosdorf; Bern 📂 Article 📅 1995 🏛 John Wiley and Sons 🌐 English ⚖ 1006 KB

## Abstract Permanent transvenous cardioverter–defibrillator leads were investigated by multiplane transesophageal echocardiography (TEE) (1) to determine whether intracardiac lead segments can be visualized, (2) to verify the position of the coils, and (3) to detect possible thrombus formation. Th