Two-channel DC-SQUID picovoltmeter for the measurement of low-field Hall coefficient and electrical resistance from 4.2 to 100 K

✍ Scribed by F Sachslehner; W Vodel


Publisher
Elsevier Science
Year
1998
Tongue
English
Weight
220 KB
Volume
38
Category
Article
ISSN
0011-2275

No coin nor oath required. For personal study only.

✦ Synopsis


A two-channel DC-SQUID voltmeter developed for the measurement of low-field Hall coefficient and electrical resistance of high-purity copper single crystals between 4.2 and 100 K is described. The noise is 1.8 pVHz -1/2 at 4.2 K and increases to 12 pVHz -1/2 at 100 K. A sample current up to 2.5 Å and a magnetic induction up to 60 mT can be applied.