Turbine flow sensor for volume-flow rate verification in MR

✍ Scribed by Richard Frayne; David W. Holdsworth; Robert F. Smith; Reza Kasrai; Jan P. T. Larsen; Brian K. Rutt


Publisher
John Wiley and Sons
Year
1994
Tongue
English
Weight
689 KB
Volume
32
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

A turbine flow sensor for MR flow experiments has been evalusted using reference volume‐flow rate measurements obtained using an electromagnetic (EM) flow meter measurements and simultaneous phase contrast (PC) MR acquisitions. After calibration, the device was found to have accuracy (compared with the EM flow meter), linearity, and precision of better than ±1%, ±3.5%, 3.5%, respectively, in constant flow mode (0 to 30 ml s^−1^). The frequency response of the flow sensor was flat (within ±10%) up to 13.9 Hz. Volume‐flow rate measurements on constant and simulated physiologic flow waveforms were in close agreement with both the electromagnetic (EM) flow meter and the gated MR PC estimates.