Tracking planar orientations of active MRI needles
✍ Scribed by Shashank Sathyanarayana; Pelin Aksit; Aravind Arepally; Parag V. Karmarkar; Meiyappan Solaiyappan; Ergin Atalar
- Publisher
- John Wiley and Sons
- Year
- 2007
- Tongue
- English
- Weight
- 431 KB
- Volume
- 26
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose
To determine and track the planar orientation of active interventional devices without using localizing RF microcoils.
Materials and Methods
An image‐based tracking method that determines a device's orientation using projection images was developed. An automated and a manual detection scheme were implemented. The method was demonstrated in an in vivo mesocaval puncture procedure in swine, which required accurate orientation of an active transvascular needle catheter.
Results
The plane of the catheter was determined using two projection images. The scan plane was adjusted automatically to follow the catheter plane, and its orientation with respect to a previously acquired target plane was displayed. The algorithm facilitated navigation for a fast and accurate puncture.
Conclusion
Using image‐based techniques, with no mechanical design changes, the orientation of an active intravascular probe could be tracked. J. Magn. Reson. Imaging 2007. © 2007 Wiley‐Liss, Inc.