Tissue strength analysis of autologous and cadaveric allografts for the pubovaginal sling

✍ Scribed by Matthew L. Lemer; David C. Chaikin; Jerry G. Blaivas


Book ID
101297838
Publisher
John Wiley and Sons
Year
1999
Tongue
English
Weight
30 KB
Volume
18
Category
Article
ISSN
0733-2467

No coin nor oath required. For personal study only.

✦ Synopsis


The aim of this study was to compare the mechanical properties of autologous rectus fascia (ARF), two groups of commercially available cadaveric fascia lata commonly used in pubovaginal sling surgery [freeze-dried (FD) and solvent-dehydrated (SD)], and commercially available cadaveric dermal grafts (DG) evaluate differences in tissue strength and stiffness. We prospectively studied the maximum load to failure (MLF) and stiffness in 20 specimens of ARF, 20 specimens of FD, 20 specimens of SD, and 10 specimens of DG. Autologous fascia was obtained from patients undergoing pubovaginal sling operation utilizing rectus fascia. Cadaveric fascia was re-hydrated in saline. All specimens were then tailored into 1 Γ— 1-cm samples and mounted onto the Instron tensiometer. Samples were loaded to failure at a 100% strain rate and force-elongation curves were generated. MLF was defined as the minimum force needed to tear the tissue. Stiffness was determined by the slope of the linear portion of the force/elongation curve between 5 and 15% strain. Statistical analysis was performed using Student's t-test. There is no statistical difference in both MLF and stiffness among ARF, SD, and DG. These data show that MLF and tissue stiffness of SD and DG are comparable to that of ARF. FD has a significantly lower MLF and is significantly less stiff than ARF, SD, and DG. The SD cadaveric fascia lata allograft and the cadaveric dermal allograft may be suitable alternatives to ARF for pubovaginal sling surgery.


πŸ“œ SIMILAR VOLUMES