## Abstract ## Purpose To assess sensitivity encoding (SENSE) for contrast‐enhanced MR angiography (CE‐MRA) of the abdominal arteries in comparison with standard MRA protocols. ## Materials and Methods In 22 patients MRA of the abdominal arteries was performed twice (once using a standard protoc
Time-resolved contrast-enhanced pulmonary MR angiography using sensitivity encoding (SENSE)
✍ Scribed by Yoshiharu Ohno; Hideaki Kawamitsu; Takanori Higashino; Daisuke Takenaka; Hirokazu Watanabe; Marc van Cauteren; Masahiko Fujii; Hiroto Hatabu; Kazuro Sugimura
- Publisher
- John Wiley and Sons
- Year
- 2003
- Tongue
- English
- Weight
- 388 KB
- Volume
- 17
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Purpose:
To evaluate the relationship between gadolinium concentration and signal-to-noise ratio (snr) on sensitivity encoding (sense) images, and determine the appropriate bolus injection protocol for visualizing pulmonary circulation.
Materials and methods:
Eighteen different gadolinium concentration phantoms (0, 0.01, 0.05, 0.1, 0.2, 0.4, 0.6, 0.8, 1.0, 1.2, 1.5, 2.0, 5.0, 10.0, 20.0, 30.0, 50.0, and 100.0 mmol/l) were analyzed to determine the relationship between gadolinium concentration and snr on sense images in a phantom study. in an in vivo study, 3 ml (protocol a) or 6 ml (protocol b) of gd-dtpa bma at 3 ml/second, and 5 ml of gd-dtpa bma at 5 ml/second (protocol c) were administered to eight normal volunteers for contrast-enhanced (ce) pulmonary mr angiography (mra) with sense. the peak snrs of pulmonary parenchyma and the difference in snr between pulmonary artery (pa) and pulmonary vein (pv) at peak snr in the pa were statistically evaluated.
Results:
For each flip angle at each gadolinium concentration, the snrs and contrast-to-noise ratios (cnrs) of the sense images were significantly lower than those acquired with a nonparallel imaging technique (p < 0.05). the peak snr of the pulmonary parenchyma, and differences in snr between the pa and pv at the peak snr of the pa obtained with a 5-ml/second bolus injection protocol were found to be significantly higher than those obtained with other protocols (p < 0.05).
Conclusion:
3d-ce-mra using sense demonstrated linearity between gadolinium concentration and snr, and resulted in mra with high spatial and temporal resolution with the aid of a sharp bolus injection protocol.
📜 SIMILAR VOLUMES
## Abstract Both high spatial resolution and high temporal resolution are desirable for contrast‐enhanced magnetic resonance angiography (CE‐MRA) in order to depict the arterial vasculature. In this work a fast MR pulse sequence (spiral time‐resolved imaging with contrast kinetics (Spiral‐TRICKS))
## Abstract ## Purpose To investigate if the use of parallel imaging is feasible and beneficial for peripheral contrast‐enhanced magnetic resonance angiography (CE‐MRA). ## Materials and Methods A total of 19 consecutive patients underwent peripheral CE‐MRA using SENSE with two‐fold reduction in
## Abstract ## Purpose To investigate the feasibility of using time‐of‐flight (TOF) images as a constraint in the reconstruction of a series of highly undersampled time‐resolved contrast‐enhanced MR images (HYPR TOF), to allow simultaneously high temporal and spatial resolution and increased SNR.
## Abstract ## Purpose To compare contrast characteristics and image quality of 1.0 M gadobutrol with 0.5 M Gd‐DTPA for time‐resolved three‐dimensional pulmonary magnetic resonance angiography (MRA). ## Materials and Methods Thirty‐one patients and five healthy volunteers were examined with a co
## Abstract ## Purpose To prospectively assess the influence of a gadolinium doped saline flush compared with a pure saline flush on the image quality of the supra‐aortic vessels using time‐resolved contrast‐enhanced MR angiography (4D CE‐MRA) in a randomized double blind clinical trial. ## Mater