𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Thermal effects of stationary “hot tip” laser coronary probes: An in vitro assessment

✍ Scribed by Dr. Eric Rosenthal; Dr. Joseph K. Montarello; Tim Palmer; Paul V. L. Curry


Publisher
John Wiley and Sons
Year
1989
Tongue
English
Weight
751 KB
Volume
9
Category
Article
ISSN
0196-8092

No coin nor oath required. For personal study only.

✦ Synopsis


Percutaneous laser thermal probe angioplasty requires sufficient laser probe flexibility to access the coronary tree. This may entail a loss of axial strength and the resultant slow advancement may lead to unwanted heating of the normal coronary artery proximal to the lesion. To assess the lateral thermal effects of stationary coronary laser probes, laser thermal energy (50-150 J) was delivered to 25 coronary artery segments (diameter 1 . 9 4 0 mm) in a perfused cadaver heart preparation using a 1.7 mm tip probe. Adherence to the vessel wall occurred in 19 segments, endothelial charring in 8 segments, and perforation in 3 segments. Endothelial charring was seen in 8 of 13 nonperfused segments but in 0 of 12 segments perfused at 60 ml/minute (P < 0.01).

In all three perforations the vessel to probe diameter ratio was less than 1.6:1, perfusion was absent, and traction to dislodge the adherent probe was necessary. Lateral wall damage is a complication of stationary laser probes: smaller-tipped probes which are advanced rapidly at the time of energy delivery may enhance the safety margins of coronary laser thermal probe angioplasty.


📜 SIMILAR VOLUMES


Thermal effects of CO2 laser on the pulp
✍ Anna Marie A. Arrastia; Petra Wilder-Smith; Michael W. Berns 📂 Article 📅 1995 🏛 John Wiley and Sons 🌐 English ⚖ 632 KB

Background and Objective: Thermal effects of direct exposure of the pulp chamber to C 0 2 laser irradiation were evaluated and potentially safe parameters for its clinical application were determined. Study DesignfMaterials and Methods: Using conventional methods, the pulp chambers of freshly extrac