The testing and adjustment of Lyot-Öhman filters

✍ Scribed by J. V. Ramsay; D. G. Norton; E. G. V. Mugridge


Book ID
104642011
Publisher
Springer
Year
1968
Tongue
English
Weight
122 KB
Volume
4
Category
Article
ISSN
0038-0938

No coin nor oath required. For personal study only.

✦ Synopsis


To determine the line profile of a Lyot-Ohman filter, an instrument with a spectral resolution greater than that of the filter is required, usually a grating spectrograph ( VAN GRmTHUYS~N and HOUTGAST, 1959). Alternatively, we describe below the use of an automatically controlled scanning Fabry-Perot interferometer (RAMSAY, 1966) for this purpose.

The plates of this interferometer are mounted on rods of the piezoelectric material barium titanate, the parallelism being controlled automatically independent of plate spacing using white-fight superposition fringes for testing purposes. In addition, the interferometer is equipped with a small, dimensionally stable wedge-shaped calibrated reference interferometer. The main interferometer can similarly be controlled to transmit the same wavelength as a selected aperture of this calibrated reference interferometer. Alternatively, if the interferometer is used in a scanning mode, the superposition fringes between the reference interferometer and the main working interferometer may be used for wavelength calibration during scanning. Further details of the interferometer are available in RAMSAY (1966).

We have used this instrument in the scanning mode when testing birefringent filters. The filter is illuminated by collimated light from a Pointolite source and the transmitted radiation is scanned by the interferometer. The spectral content of the transmitted radiation (the line profile) is displayed on one channel of a two-pen recorder whilst the superposition fringes between the reference and working interferometers is displayed on the other channel. A typical result is shown in Figure 1. The overall operation is simple and rapid: the interpretation of the results is very convenient and the line profile can be plotted immediately an adjustment is made. The resolution of the interferometer may readily be chosen to suit the bandwidth of the filter under test.

The method we have used to adjust our filters is as follows. Suppose the filter has two tunable elements, A and B, and let B have the higher resolving power. Effectively take B out of the system (by removing its appropriate polaroid), then rotate the polaroid which tunes A until the two pass-bands which appear are of the same amplitude. Note the orientation of the polaroid. Then rotate the polaroid approximately 180 ~ and repeat. Half-way between these two orientations is the position to which element A should be adjusted for synchronism with the preceding (lower resolution) elements. Bring element B into the system and rotate polaroid B until the side bands are again equal, repeating with B rotated through approximately 180 ~ Polaroid B