๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

The minimum crystal size needed for a complete diffraction data set

โœ Scribed by Holton, James M. ;Frankel, Kenneth A.


Book ID
104478542
Publisher
International Union of Crystallography
Year
2010
Tongue
English
Weight
840 KB
Volume
66
Category
Article
ISSN
0907-4449

No coin nor oath required. For personal study only.

โœฆ Synopsis


In this work, classic intensity formulae were united with an empirical spot-fading model in order to calculate the diameter of a spherical crystal that will scatter the required number of photons per spot at a desired resolution over the radiation-damage-limited lifetime. The influences of molecular weight, solvent content, Wilson B factor, X-ray wavelength and attenuation on scattering power and dose were all included. Taking the net photon count in a spot as the only source of noise, a complete data set with a signal-to-noise ratio of 2 at 2โ€…ร… resolution was predicted to be attainable from a perfect lysozyme crystal sphere 1.2โ€…ยตm in diameter and two different models of photoelectron escape reduced this to 0.5 or 0.34โ€…ยตm. These represent 15-fold to 700-fold less scattering power than the smallest experimentally determined crystal size to date, but the gap was shown to be consistent with the background scattering level of the relevant experiment. These results suggest that reduction of background photons and diffraction spot size on the detector are the principal paths to improving crystallographic data quality beyond current limits.