The impact of partial-volume effects in dynamic susceptibility contrast magnetic resonance perfusion imaging
✍ Scribed by Jean J. Chen; Michael R. Smith; Richard Frayne
- Publisher
- John Wiley and Sons
- Year
- 2005
- Tongue
- English
- Weight
- 869 KB
- Volume
- 22
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose
To demonstrate the degree of the cerebral blood flow (CBF) estimation bias that could arise from distortion of the arterial input function (AIF) as a result of partial‐volume effects (PVEs) in dynamic susceptibility contrast (DSC) magnetic resonance imaging (MRI).
Materials and Methods
A model of the volume fraction an artery occupies in a voxel was devised, and a mathematical relationship between the amount of PVE and the measured baseline MR signal intensity was derived. Based on this model, simulation studies were performed to assess the impact of PVE on CBF. Furthermore, the effectiveness of linear PVE compensation approaches on the concentration function was investigated.
Results
Simulation results showed a nonlinear relationship between PVE and the resulting CBF measurement error. In addition to AIF underestimation, PVE also causes distortions of AIF frequency characteristics, leading to CBF errors varying with mean transit time (MTT). An uncorrected AIF measured at a voxel with a partial‐volume fraction of ≤50% could produce a CBF overestimation of more than fourfold. Linear compensation of the concentration curves did not produce correct CBF estimates.
Conclusion
PVE can induce significant CBF estimation biases. In addition, the MTT dependence of CBF accuracy raises doubts of the validity of adopting a single cross‐calibration factor (i.e., setting normal white matter to 22 mL minute^−1^ (100 g)^−1^) to obtain CBF values with absolute units. The impact of PVE may be reduced by decreasing the maximum arterial signal drop in the perfusion images. To correct the AIF distortions introduced by PVE, the nonlinear relationship between the impact of PVE on MR signal intensity and contrast concentration function must be considered. J. Magn. Reson. Imaging 2005. © 2005 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract ## Purpose To identify radiation‐induced changes in the cerebral vasculature of healthy tissue in the first four months following radiotherapy through the analysis of dynamic‐susceptibility contrast perfusion imaging. ## Materials and Methods Dynamic gradient‐echo imaging was perform
The quantitation of in vivo 1 H MR spectroscopy and dynamic contrast enhanced MR imaging is described for patients with histologically confirmed prostate adenocarcinoma and benign prostatic hypertrophy (BPH). Results are presented which suggest that combined use of these techniques may be helpful in
## Abstract ## Purpose To evaluate the feasibility of using dynamic contrast‐enhanced magnetic resonance imaging (MRI) for assessment of muscle perfusion in a rat model of hind‐limb ischemia. ## Materials and Methods The acute alteration and chronic recovery in muscle perfusion and perfusion res
## Abstract Magnetic resonance imaging enhanced with a macromolecular contrast medium (MMCM), albumin‐Gd‐DTPA, was used to estimate the plasma volume __in vivo__ in the myocardium, lung, liver, and skeletal muscle of 10 normal rats. The plasma volumes of the same tissues in a parallel group of six