The Emit® 2000 Cyclosporine Specific Assay, Extended Range: Development of an application protocol for the V-Twin® Analyzer

✍ Scribed by Nihmat Morjana; Harriet Yau; Diane Rea; Doris Ruttle; Heather Jones; Gerald Siefring Jr.; Robert Christenson


Publisher
Portland Press
Year
2011
Tongue
English
Weight
265 KB
Volume
58
Category
Article
ISSN
0885-4513

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

We evaluated a new protocol for measurement of cyclosporine A (CsA) 2 H after dose (C2) on the V‐Twin® analyzer. Imprecision, recovery, and linearity were determined using CsA‐spiked blood pools. Accuracy was evaluated using specimens from renal, cardiac, and liver transplant patients, and results were compared with those from liquid chromatography–tandem mass spectrometry (LC–MS/MS) and the Abbott TDx®/TDxFLx® assay. Cross‐reactivity and interferences were assessed in the presence of 800 ng/mL CsA. Imprecision coefficients of variation were 3.3%–4.8% (within run) and 5.9%–8.7% (total). Recovery was within 10% of the expected values. Linearity was 350–2,000 ng/mL. Calibration was stable for ≥ 2 weeks. Method comparison showed regression statistics: V‐Twin® = 1.01 × LC tandem MS + 36.1, r = 0.971; V‐Twin® = 1.13 × Abbott − 92.4, r = 0.969. Metabolite cross‐reactivity and interference (endogenous substances and drugs) were within ±10%. The C2 protocol on the V‐Twin® analyzer provides acceptable assay performance and accurate determination of whole blood CsA drawn at 2 H after dose.