## Abstract Synthetic calcium phosphate ceramic (CPC) surfaces can be transformed to a biological apatite through a sequence of reactions which include dissolution, precipitation, and ion exchange. By virtue of the reactions being materialโdependent, it is important to determine parametric rate eff
The effects of lubricant composition onin vitro wear testing of polymeric acetabular components
โ Scribed by Wang, A. ;Essner, A. ;Schmidig, G.
- Publisher
- John Wiley and Sons
- Year
- 2003
- Tongue
- English
- Weight
- 156 KB
- Volume
- 68B
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
โฆ Synopsis
Abstract
Lubricant composition is known to affect in vitro (simulator) wear of polymeric acetabular components. Clinical acetabular components, fabricated from both ultrahigh molecular weight polyethylene (UHMWPE) and polytetrafluoroethylene (PTFE), were tested against cobaltโchromium 32โmm ball heads, with the use of various lubricants, with different compositions and concentrations of bovine and calf serum as well as hyaluronic acid, in tests which lasted one million cycles. The type of proteins and their relative concentration in the lubricant affected the wear rate of both UHMWPE and PTFE. Increasing protein concentration reduced the UHMWPE wear rate but increased the PTFE wear rate. For a given total protein concentration, increasing the albumin/globulin (A/G) ratio led to a significant reduction in the wear rate for PTFE. A similar but smaller effect was found for UHMWPE. The relative wearโrate ratio between PTFE and UHMWPE depends on both the total protein concentration and the specific protein concentration (albumin/globulin ratio). The average clinical wear rates for both PTFE and UHMWPE and the average wearโrate ratios between them were reproduced when the serum lubricant contained total protein concentration and A/G ratio within normally observed physiological ranges. It is recommended that lubricants for simulator testing be standardized to control total protein content as well as albumin/globulin ratio. ยฉ 2003 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 68B: 45โ52, 2004
๐ SIMILAR VOLUMES
The formation of a biologically equivalent carbonatecontaining apatite on the surface of synthetic calcium phosphate ceramics (CPC) may be an important step leading to bonding with bone. Reactions of several single phases CPCs upon immersion into a simulated physiologic solution (SPS) with an electr
The effects of lubricating-oil additive zinc dialkyldithiophosphate (ZDDP) on the friction and wear properties of polymers and their composites sliding against GCr15 bearing steel were studied by using an MHK-500 ring-on-block wear tester (Timken wear tester). Then the frictional surfaces of the fri