โœฆ Synopsis


Abstract

Silicone spacers have been in use as replacement joints in the human hand for over 30 years. Since they were first used there has been a number of designs all of which have had problems with fracture. This may be due to a defect in the material caused during implantation, or by bony intrusions within the arthritic hand after implantation. The aim of this research was to investigate the effect of the environment on the mechanical properties of medical grade silicones used for human implantation. The materials were subjected to static tensile testing after various forms of ageing. The environmental conditions included temperatures of 37 and 80ยฐC and the environments of Ringer's solution, distilled water, and air. The environmental conditions employed resulted in reduced mechanical strength with ageing time of the silicones. This research supports the view that failure of silicone implants in the hand could be partly attributed to the effects of environmental ageing of the material. ยฉ 2008 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 2008


๐Ÿ“œ SIMILAR VOLUMES


Response to letter to the editor: The ef
โœ Laura J. Leslie; Mike J. Jenkins; Duncan E. T. Shepherd; Stephen N. Kukureka ๐Ÿ“‚ Article ๐Ÿ“… 2009 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 33 KB

The intention of the article 1 was to assess the likely impact of the environment on medical-grade silicone, and what effect this may have on the material properties. Two medical-grade silicones were chosen for their similarity to those materials used previously as small joint replacements. The diff

Effect of saline exposure on the surface
โœ Kennan, J. J. ;Peters, Y. A. ;Swarthout, D. E. ;Owen, M. J. ;Namkanisorn, A. ;Ch ๐Ÿ“‚ Article ๐Ÿ“… 1997 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 261 KB ๐Ÿ‘ 2 views

Medical-grade silicone elastomers were subjected to accelerated aging in saline to verify the hydrolytic stability of the elastomer. Tensile strength, elongation at break, and the elastomer stress measured at 100% or 200% elongation did not change significantly for peroxide-cured sheeting aged in 37

Mechanical properties of medical grade e
โœ Catanese, J. ;Cooke, D. ;Maas, C. ;Pruitt, L. ๐Ÿ“‚ Article ๐Ÿ“… 1999 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 233 KB ๐Ÿ‘ 2 views

Expanded polytetrafluoroethylene (e-PTFE) is used successfully in a multitude of biomedical and clinical applications. The success of this biomaterial is due to its microporous structure that allows biointegration for fixation, as well as overall mechanical integrity. The mechanical properties and d