The effect of single use laryngoscopy equipment on illumination for tracheal intubation
β Scribed by K. J. Anderson; N. Bhandal
- Publisher
- John Wiley and Sons
- Year
- 2002
- Tongue
- English
- Weight
- 80 KB
- Volume
- 57
- Category
- Article
- ISSN
- 0003-2409
No coin nor oath required. For personal study only.
β¦ Synopsis
Summary We measured the illumination generated by all 30 Macintosh size 3 laryngoscopes in our department with a luxβmeter and a standardised laryngoscope holding tube. We found a large range in illumination generated (65β3130 lx). We then measured the effect on the illumination for each laryngoscope by covering the blade with a cover (LaryGard). In every case, the illumination was reduced by the LaryGard, the mean (SD) reduction was 67 (19)%. When we compared the illumination generated by a disposable laryngoscope blade with the same power source, we found that the illumination was reduced less than with the standard Macintosh covered by a LaryGard.