๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

The bovine pericardial xenograft: III. Effect of uniaxial and sequential biaxial stress during fixation on the tensile viscoelastic properties of bovine pericardium

โœ Scribed by Lee, J. Michael ;Ku, Michael ;Haberer, Sean A.


Publisher
John Wiley and Sons
Year
1989
Tongue
English
Weight
919 KB
Volume
23
Category
Article
ISSN
0021-9304

No coin nor oath required. For personal study only.

โœฆ Synopsis


Our previous two articles have shown that glutaraldehyde-fixed bovine pericardium is nearly isotropic, whether fixed without constraints, with tethering, or with pressure. In this study, we have used uniaxial stress during fixation to produce bovine pericardial material with marked tensile anisotropy. Rectangular and cruciate pericardial samples have been mechanically examined after one of four treatments: (i) fixation under 88-kPa uniaxial stress, (ii) fixation under 176-kPa uniaxial stress, (iii) 3 h of 176-kPa uniaxial stress in saline followed by 24-h fixation under the same stress, (iv) fixation under 176-kPa uniaxial stress followed by a second fixation under 176-kPa stress in a direction normal to the first. Strips of material were cut at O", 30", 60ยฐ, and 90" to direction of the initial stress, and tested for response to cyclic loading, stress relaxation, plastic deformation, and fracture properties. Fixation under uniaxial stress produced anistropy similar to that seen in porcine aortic valve leaflets; however, the overall extensibility of the material depended on the applied stress and the aspect ratio of the stressed sample. While loading in saline produced no change, the sequential biaxial stressing produced a reduction in anisotropy, suggesting exposure of additional crosslinking sites. Uniaxial stress during fixation may be a useful method for construction of anisotropic heart valve leaflets.


๐Ÿ“œ SIMILAR VOLUMES


The bovine pericardial xenograft: II. Ef
โœ Lee, J. Michael ;Corrente, Ralph ;Haberer, Sean A. ๐Ÿ“‚ Article ๐Ÿ“… 1989 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 744 KB

Our previous article suggested that control of the extensibility of aldehyde-fixed pericardium could be achieved by controlling shrinkage during fixation. Therefore, to prevent shrinkage, we have used sandpaper-lined plexiglass plates to clamp circular samples of bovine pericardium during fixation i